Light, versatile, and easy to use, Trilogy makes invasive and noninvasive home ventilation less complicated with a simplified user interface and Respironics-proven BiPAP technology for greater versatility. Its lightweight design and internal and detachable batteries make portability easier for patients on the go. Whether at home or in alternative care sites, the Trilogy 100 portable volume and pressure support ventilator with BiPAP technology is made for adult and pediatric use.
Respironics Trilogy 200 Ventilator,
Environmental
Temperature 5°C – 40°C
Relative humidity 15% – 95%
Atmospheric pressure 110 kPa to 60 kPa
Physical
Weight 11 lbs, 5 kgs
Size 6.6"L x 11.2"W x 9.3"H (16.68 cm L x 28.45 cm W x 23.52 cm H)
Controls
Circuit types Active with PAP, passive
Ventilation types Volume control, pressure control, bilevel
Volume modes AC, SIMV (w/PS), CV
Pressure modes CPAP, S, S/T, T, PC-SIMV (w/PS)
IPAP 4 – 50 cm H20
EPAP 0 – 25 cm H20 for active circuits
4 – 25 cm H20 for passive circuits
CPAP 4 – 20 cm H20 for passive circuits
PEEP 0 – 25 cm H20 for active circuits
4 – 25 cm H20 for passive circuits
Pressure support differential 0 – 30 cm H20
Measured patient parameters
Tidal volume 50 – 2000 ml
Breath rate 0 – 60 for AC mode
1 – 60 for all other modes
Inspiratory time 0.3 – 5.0 secs
Rise time 1 – 6
Ramp start pressure 0 – 25 cm H20 for active circuits
4 – 25 cm H20 for passive circuits
4 – 19 cm H20 in CPAP mode
Ramp length Off, 5 – 45 minutes
Flex Off, 1 – 3
Flow trigger sensitivity 1 – 9 l/min
Flow cycle 10% – 90%
Apnea rate 4 – 60 BPM
Tidal volume 0 – 2000 ml
Minute ventilation 0 – 99 l/min
Leak rate 0 – 200 l/min
Respiratory rate 0 – 80 BPM
Peak inspiratory flow 0 – 200 l/min
Peak inspiratory pressure 0 – 99 cm H20
Mean airway pressure 0 – 99 cm H20
Percentage patient triggered breaths 0 – 100%
I:E ratio 9.9:1 to 1:9.9
Alarms
Circuit disconnect Off, 10 – 60 secs
Apnea Off, 10 – 60 secs
High tidal volume Off, 50 – 2000 ml
Low tidal volume Off, 50 – 2000 ml
High minute ventilation Off, .1 – 99 l/min
Low minute ventilation Off, .1 – 99 l/min
High respiratory rate Off, 4 – 80 BPM
Low respiratory rate Off, 4 – 80 BPM