Trilogy provides respiratory support to meet your prescribed needs, up to 24 hours a day. Trilogy is portable with internal, hot-swappable detachable, and external batteries to provide greater freedom and mobility with continuous ventilation.

Description

Simple Ventilation. Sophisticated Technology. Trilogy makes invasive and noninvasive home ventilation less complicated with a simplified user interface and Respironics' proven BiPAP technology for greater versatility. Its lightweight design and internal and detachable batteries make portability easier for patients on the go. Simplicity Light, versatile, and easy to use, Trilogy100 offers clinicians and patients one of life's most remarkable qualities - simplicity. Whether at home or in alternative care sites, the Trilogy100 portable volume and pressure support ventilator with BiPAP technology is made for adult and pediatric use. Offers Two Circuit Options Interchangeable active and passive exhalation porting blocks give clinicians the flexibility to choose the best available circuit and interface for patients. Its internal PEEP, combined with a single limb circuit, creates an uncomplicated application. Portability Trilogy100 has been designed with portability in mind. • Weighs 5 kg (11 lb.) • Includes 3-hour internal and 3-hour detachable batteries • Ability to connect to an external lead-acid battery DirectView Ventilation Management Software Trilogy100 provides ventilation data and prescription management through the use of an SD memory card.

Environmental Specifications

  • Temperature: 5°C – 40°C
  • Relative humidity: 15% – 95%
  • Atmospheric pressure: 110 kPa to 60 kPa

Physical Specifications

  • Weight: 11 lbs, 5 kgs
  • Size: 6.6"L x 11.2"W x 9.3"H (16.68 cm L x 28.45 cm W x 23.52 cm H)

Controls

  • Circuit types: Active with PAP, passive
  • Ventilation types: Volume control, pressure control, bilevel
  • Volume modes: AC, SIMV (w/PS), CV
  • Pressure modes: CPAP, S, S/T, T, PC-SIMV (w/PS)
  • IPAP: 4 – 50 cm H20
  • EPAP: 0 – 25 cm H20 for active circuits; 4 – 25 cm H20 for passive circuits
  • CPAP: 4 – 20 cm H20 for passive circuits
  • PEEP: 0 – 25 cm H20 for active circuits; 4 – 25 cm H20 for passive circuits
  • Pressure support differential: 0 – 30 cm H20

Measured Patient Parameters

  • Tidal volume: 50 – 2000 ml
  • Breath rate: 0 – 60 for AC mode, 1 – 60 for all other modes
  • Inspiratory time: 0.3 – 5.0 secs
  • Rise time: 1 – 6
  • Ramp start pressure: 0 – 25 cm H20 for active circuits; 4 – 25 cm H20 for passive circuits; 4 – 19 cm H20 in CPAP mode
  • Ramp length: Off, 5 – 45 minutes
  • Flex: Off, 1 – 3
  • Flow trigger sensitivity: 1 – 9 l/min
  • Flow cycle: 10% – 90%
  • Apnea rate: 4 – 60 BPM
  • Tidal volume: 0 – 2000 ml
  • Minute ventilation: 0 – 99 l/min
  • Leak rate: 0 – 200 l/min
  • Respiratory rate: 0 – 80 BPM
  • Peak inspiratory flow: 0 – 200 l/min
  • Peak inspiratory pressure: 0 – 99 cm H20
  • Mean airway pressure: 0 – 99 cm H20
  • Percentage patient triggered breaths: 0 – 100%
  • I:E ratio: 9.9:1 to 1:9.9

Alarms

  • Circuit disconnect: Off, 10 – 60 secs
  • Apnea: Off, 10 – 60 secs
  • High tidal volume: Off, 50 – 2000 ml
  • Low tidal volume: Off, 50 – 2000 ml
  • High minute ventilation: Off, .1 – 99 l/min
  • Low minute ventilation: Off, .1 – 99 l/min
  • High respiratory rate: Off, 4 – 80 BPM
  • Low respiratory rate: Off, 4 – 80 BPM